Abstract

Objective: Compare vascular complications and incidence of bleeding of Impella 2.5 and intra-aortic balloon pump (IABP) in high-risk percutaneous coronary interventions (PCI).

Background: Large arterial sheath size for device insertion is associated with vascular and/or bleeding complications; gastrointestinal bleeding may also occur with anti-coagulation use.

Methods: Patients with an acute coronary syndrome receiving Impella 2.5 or IABP during high-risk PCI were studied (13 Impella; 62 IABP). Vascular complications and incidence of bleeding were compared.

Results: Post-procedure hematocrit was similar between groups. Blood transfusion occurred in 38.4% and 32.2% of patients in the Impella and IABP groups, respectively (P = NS); 65.3%, 30.7% and 3.8% of bleeding were due to vascular access site/procedure related, gastrointestinal and genitourinary, respectively. There was no statistical significant difference in vascular complications between the Impella and IABP groups (15.3% and 6.4% of patients, respectively); mesenteric ischemia (n = 1) and aortic rupture (n = 1) were only in the IABP group. In-hospital and one-year mortality were not statistically significant between groups.

Conclusion: Impella can be used as safely as IABP during high-risk PCI with similar vascular and bleeding complications. Importantly, approximately one third of bleeding was from the gastrointestinal system warranting careful prophylactic measures and monitoring.

Publication types

  • Comparative Study

MeSH terms

  • Acute Coronary Syndrome / mortality
  • Acute Coronary Syndrome / therapy*
  • Aged
  • Blood Transfusion / statistics & numerical data
  • Coronary Stenosis
  • Female
  • Gastrointestinal Hemorrhage / etiology
  • Heart-Assist Devices / adverse effects*
  • Hematocrit
  • Hemodynamics
  • Hemorrhage / etiology*
  • Hospital Mortality
  • Humans
  • Intra-Aortic Balloon Pumping / adverse effects*
  • Male
  • Middle Aged
  • Retrospective Studies