Preclinical investigation of ab interno trabeculectomy using a novel dual-blade device

Am J Ophthalmol. 2013 Mar;155(3):524-529.e2. doi: 10.1016/j.ajo.2012.09.023. Epub 2012 Dec 4.

Abstract

Purpose: To evaluate the effects of a novel ab interno trabeculectomy device on human trabecular meshwork (TM).

Design: Laboratory evaluation.

Methods: The TM from human cadaveric corneal rim tissue was incised using 3 instruments: (1) novel dual-blade device; (2) microvitreoretinal (MVR) blade; and (3) Trabectome. Tissue samples underwent histologic processing and comparative analyses. Subsequently, human eye perfusion studies were performed to evaluate intraocular pressure (IOP)-lowering effects of each device. Main outcome measures were degree of TM removal by histology and IOP in a perfusion model.

Results: The MVR blade exhibited minimal removal of TM and obvious injury to the adjacent sclera. The Trabectome removed a large portion of the central TM, but leaflets of residual tissue remained and thermal injury was noted in all samples. The dual-blade device achieved a more complete removal of TM without injury to surrounding tissues. All devices resulted in statistically significant lowering of IOP during perfusion model studies. MVR blade treatment across 170.0 ± 14.1 degrees of TM resulted in a decrease of IOP from 18.5 ± 1.9 mm Hg to 12.8 ± 2.2 mm Hg (P < .01). Trabectome treatment across 117.5 ± 12.6 degrees resulted in a decrease of IOP from 18.8 ± 1.7 mm Hg to 11.3 ± 1.0 mm Hg (P < .01). Dual-blade device treatment across 157.5 ± 26.3 degrees resulted in a decrease of IOP from 18.3 ± 3.0 mm Hg to 11.0 ± 2.2 mm Hg (P < .01).

Conclusions: The novel dual-blade device demonstrated a more complete removal of TM without residual TM leaflets or damage to surrounding tissues and significantly reduced IOP in a human eye perfusion model.

Publication types

  • Evaluation Study

MeSH terms

  • Humans
  • Intraocular Pressure
  • Tissue Donors
  • Trabecular Meshwork / surgery*
  • Trabeculectomy / instrumentation*