Comparison of plastic and metallic single-use and metallic reusable laryngoscope blades: a randomised controlled trial

Eur J Anaesthesiol. 2013 Apr;30(4):163-9. doi: 10.1097/EJA.0b013e32835c1cdb.

Abstract

Context: Plastic single-use laryngoscope blades have been found to increase the difficulty of intubation. Metallic single-use blades may represent an alternative.

Objective: To compare the rates of intubation failure at first laryngoscopy using metallic single-use, plastic single-use and metallic reusable laryngoscope blades.

Design: Randomised, superiority, single-centre, controlled trial.

Setting: Operating theatres at a University Hospital from January 2008 to August 2009.

Patients: A total of 1863 adults requiring general anaesthesia.

Intervention: Patients were randomised to one of three laryngoscope blades: metallic single-use, plastic single-use or metallic reusable.

Main outcome measure: The primary end-point was the number of intubation failures at first laryngoscopy. Secondary end-points were glottic exposure and the Intubation Difficulty Scale.

Results: The failure rate (absolute difference: 95% confidence interval) at the first attempt was greater for the plastic single-use blades (8.1%) than for the metallic reusable [3.2% (4.9%: 2.2 to 7.6%); P < 0.001] and metallic reusable blades [4.0% (4.0%: 1.2 to 6.9%); P < 0.006]. No difference was found between the metallic reusable and metallic single-use groups (0.9%: -1.3 to 3.1%). Scoring on the Intubation Difficulty Scale [median (interquartile range)] was higher in the plastic single-use group [1 (0 to 2)] than in the metallic groups [metallic reusable: 0 (0 to 1); P < 0.001 and metallic single-use: 0 (0 to 1); P < 0.007] groups. Glottic exposure was significantly better in the metallic reusable group (modified Cormack & Lehane score III and IV: 3.7%) compared with the plastic single-use group (modified Cormack & Lehane score III and IV: 9.4%; P < 0.03).

Conclusion: The rates of failed intubation at first laryngoscopy were similar in the metallic reusable and metallic reusable groups, but greater in the plastic single-use group.

Trial registration: ClinicalTrials.gov NCT00876967.

Publication types

  • Comparative Study
  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Anesthesia, General / methods
  • Disposable Equipment
  • Equipment Design
  • Equipment Reuse
  • Female
  • Glottis
  • Hospitals, University
  • Humans
  • Intubation, Intratracheal / instrumentation*
  • Intubation, Intratracheal / methods
  • Laryngoscopes*
  • Laryngoscopy / instrumentation*
  • Male
  • Metals
  • Middle Aged
  • Plastics
  • Prospective Studies

Substances

  • Metals
  • Plastics

Associated data

  • ClinicalTrials.gov/NCT00876967