Bolus timing in high-pitch CT angiography of the aorta

Eur J Radiol. 2013 Jun;82(6):1028-33. doi: 10.1016/j.ejrad.2013.01.004. Epub 2013 Jan 31.

Abstract

Objective: To investigate the bolus geometry in high-pitch CT angiography (CTA) of the aorta without ECG synchronisation in comparison to single-source CT.

Methods: Overall 160 consecutive patients underwent CTA either in conventional single-source mode with a pitch of 1.2 (group 1), or in dual-source mode with a pitch of 3.0 (groups 2, 3 and 4) using different contrast media timings with bolus triggering at 140 HU (5s, group 1; 10s, group 2; 12s, group 3; 14s, group 4). Contrast material, saline flush, flow rate and kV/mAs settings were kept equal for optimum comparability. Aortic attenuation was measured along the z-axis of the patient at different anatomic landmarks and subjective image quality was compared.

Results: The most homogeneous enhancement of the aorta was reached with a delay of 10s after reaching the trigger threshold. The imaging length was not significantly different, but the examination time was significantly (p<0.001) shorter in the high-pitch group (7.7s vs. 1.7s for group 1 vs. 2, 3 and 4).

Conclusion: In high-pitch CT angiography using a start delay of 10s after a trigger threshold of 140 HU in the descending aorta is reached, a homogenous contrast along the z-axis is accomplished.

Publication types

  • Comparative Study
  • Controlled Clinical Trial

MeSH terms

  • Adult
  • Aged
  • Aortic Diseases / diagnostic imaging*
  • Aortography / methods*
  • Contrast Media / administration & dosage
  • Female
  • Humans
  • Male
  • Middle Aged
  • Radiographic Image Enhancement / methods
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Single-Blind Method
  • Tomography, X-Ray Computed / methods*
  • Young Adult

Substances

  • Contrast Media