Kirkpatrick-Baez microscope for hard X-ray imaging of fast ignition experiments

Rev Sci Instrum. 2013 Feb;84(2):023704. doi: 10.1063/1.4776670.

Abstract

A Kirkpatrick-Baez X-ray microscope has been developed for use on the Titan laser facility at the Lawrence Livermore National Laboratory in Fast Ignition experiments. It was developed as a broadband alternative to narrow band Bragg crystal imagers for imaging Kα emission from tracer layers. A re-entrant design is employed which allows for alignment from outside the chamber. The mirrors are coated with Pt and operate at a grazing incident angle of 0.5° providing higher resolution than an equal brightness pinhole and sufficient bandwidth to image thermally shifted characteristic Kα emission from heated Cu tracer layers in Fast Ignition experiments. The superpolished substrates (<1 Å rms roughness) had a final visible wavelength roughness of 1.7 Å after coating, and exhibited a reflectivity corresponding to an X-ray wavelength roughness of 7 ± 1 Å. A unique feature of this design is that during experiments, the unfiltered direct signal along with the one-dimensional reflections are retained on the detector in order to enable a live indication of alignment and incident angle. The broad spectral window from 4 to 9 keV enables simultaneous observation of emission from several spectral regions of interest, which has been demonstrated to be particularly useful for cone-wire targets. An experimentally measured resolution of 15 μm has been obtained at the center of the field of view.