Here we demonstrate the use of multiple Si nanochannel (NC) or nanograting (NG) instead of the conventional single nanochannel or nanowire design in biosensors. The NG devices can significantly reduce device-to-device variation, and improve device performance, e.g. higher current, higher ON/OFF ratio, smaller subthreshold slope, lower threshold voltage Vt in buffer solution. NG devices also result in higher sensor stability in buffer and diluted human serum. We believe such improvements are due to reduced discrete dopant fluctuation in the Si nanowires and biochemical noise in the solution because of the multiple-channel design. The improved devices allow us to sense pH linearly with 3-aminopropyltriethoxysilane coated devices, and to selectively detect insulin with limit of detection down to 10 fM in both buffer solution and diluted human serum without pre-purification.