Background: Validation of new affordable CD4 T cell measurement technologies is crucial specifically in resource-poor countries for antiretroviral treatment eligibility and immunologic CD4 monitoring of HIV-infected patients.
Methods: The absolute and percentage CD4 T cell counts of 258 HIV-1-infected blood samples (182 adults and 76 children), living in N'Djamena, Chad, were performed by single-platform, volumetric, CD45-assisted pan-leucogating Auto40 flow cytometer (Apogee Flow Systems Ltd, Hemel Hempstead, UK) comparing to the FACSCalibur flow cytometer as a reference method.
Results: Absolute and percentage CD4 T cell counts obtained by Auto40 and FACSCalibur of 258 HIV-1-infected blood samples were highly correlated (r = 0.99 and r = 0.96, respectively). The mean absolute bias and percent bias between Apogee Auto40 and FACSCalibur absolute CD4 T cell counts, were -9.4 cells/μl with limits of agreement from -15 to 93 cells/μl, and +2.0% with limits of agreement from -0.9 to 4.9%, respectively. The mean of absolute bias and percent bias between Apogee Auto40 and FACSCalibur of CD4 percentage results were +0.4% (95% CI: -0.02 - 0.86) with limits of agreement from -2.4 to 0.3%, and +3.0% with limits of agreement from -6.6 to 0.6%, respectively. The Auto40 counting allowed to identify the majority of adults with CD4 T cells below 200 cells/μl (sensitivity: 89%; specificity: 99%) or below 350 cells/μl (sensitivity: 94%; specificity:98%); and of children below 750 cells/μl (sensitivity: 99%; specificity: 96%) or below 25% CD4+ (sensitivity: 94%; specificity: 98%).
Conclusion: The Auto40 analyzer is an alternative flow cytometer for CD4 T lymphocyte enumeration to be used in routine for immunological monitoring according to the current WHO recommendations in HIV-infected adults as well as children living in resource-constrained settings like Chad.