MR guided applicator reconstruction for brachytherapy of cervical cancer using the novel titanium Rotterdam applicator

Radiother Oncol. 2013 Apr;107(1):88-92. doi: 10.1016/j.radonc.2013.03.014. Epub 2013 Apr 29.

Abstract

A novel model of the titanium Rotterdam tandem and ovoid applicator is presented. As titanium produces artefacts in MR images, an MR sequence was sought and optimised for visualisation and accurate applicator reconstruction. The mean inter-observer (8 observers) variability for four patients was only 0.7 mm (maximum 1.7 mm).

MeSH terms

  • Brachytherapy / instrumentation*
  • Brachytherapy / methods
  • Female
  • Humans
  • Magnetic Resonance Imaging, Interventional / methods*
  • Radiotherapy Dosage
  • Radiotherapy, Image-Guided / methods*
  • Titanium
  • Uterine Cervical Neoplasms / pathology
  • Uterine Cervical Neoplasms / radiotherapy*

Substances

  • Titanium