Multi-sensor esophageal temperature probe used during radiofrequency ablation for atrial fibrillation is associated with increased intraluminal temperature detection and increased risk of esophageal injury compared to single-sensor probe

J Cardiovasc Electrophysiol. 2013 Sep;24(9):958-64. doi: 10.1111/jce.12180. Epub 2013 Jun 7.

Abstract

Background: Radiofrequency (RF) ablation in the posterior left atrium has risk of thermal injury to the adjacent esophagus. Increased intraluminal esophageal temperature has been correlated with risk of esophageal injury. The objective of this study was to compare esophageal temperature monitoring (ETM) using a multi-sensor temperature probe with 12 sensors to a single-sensor probe during catheter ablation for atrial fibrillation (AF).

Methods and results: We compared the detection of intraluminal esophageal temperature rises in 543 patients undergoing RF ablation for AF with ETM. Esophageal endoscopy (EGD) was performed on all patients with maximum esophageal temperature ≥ 39°C. Esophageal lesions were classified by severity as mild or severe ulcerations. Four hundred fifty-five patients underwent RF ablation with single-sensor ETM and 88 patients with multi-sensor ETM. Thirty-nine percent of patients with single-sensor versus 75% with multi-sensor ETM reached a maximum detected esophageal temperature ≥ 39°C (P < 0.0001). Esophageal injury was detected by EGD in 29% of patients with maximum temperature ≥ 39°C by single-sensor versus 46% of patients with multi-sensor ETM (P = 0.021). Thirty-nine percent of patients with lesions in the single-sensor probe group had severe ulcerations compared to 33% of patients in the multi-sensor probe group (P = 0.641).

Conclusions: Intraluminal esophageal temperature ≥ 39°C is detected more frequently by the multi-sensor temperature probe versus the single-sensor probe, with more frequent esophageal injury and with comparable severity of injury. Despite detecting esophageal temperature rises in more patients, the multi-sensor probe may not have any measurable benefit compared to a single-sensor probe.

Keywords: atrial fibrillation; catheter ablation; complications; electrophysiology; esophageal injury; esophageal temperature monitoring; pulmonary vein isolation.

MeSH terms

  • Aged
  • Atrial Fibrillation / diagnosis*
  • Atrial Fibrillation / physiopathology
  • Atrial Fibrillation / surgery*
  • Body Temperature / physiology*
  • Catheter Ablation / adverse effects*
  • Catheter Ablation / instrumentation
  • Esophagus / injuries*
  • Esophagus / physiology
  • Female
  • Follow-Up Studies
  • Humans
  • Intraoperative Neurophysiological Monitoring / adverse effects*
  • Intraoperative Neurophysiological Monitoring / instrumentation
  • Male
  • Middle Aged
  • Retrospective Studies