Purpose: A multi-institution dosimetry intercomparison has been undertaken of prostate intensity modulated radiation therapy (IMRT) delivery. The dosimetry intercomparison was incorporated into the quality assurance for site credentialing for the Trans-Tasman Radiation Oncology Group Prostate Fractionated Irradiation Trial 08.01 clinical trial.
Methods: An anthropomorphic pelvic phantom with realistic anatomy was used along with multiplanar dosimetry tools for the assessment. Nineteen centers across Australia and New Zealand participated in the study.
Results: In comparing planned versus measured dose to the target at the isocenter within the phantom, all centers were able to achieve a total delivered dose within 3% of planned dose. In multiplanar analysis with radiochromic film using the gamma analysis method to compare delivered and planned dose, pass rates for a 5%/3 mm criterion were better than 90% for a coronal slice through the isocenter. Pass rates for an off-axis coronal slice were also better than 90% except for one instance with 84% pass rate.
Conclusions: Strengths of the dosimetry assessment procedure included the true anthropomorphic nature of the phantom used, the involvement of an expert from the reference center in carrying out the assessment at every site, and the ability of the assessment to detect and resolve dosimetry discrepancies.