Objective: To compare the outcomes of flexible ureterorenoscopy and lasertripsy (FURS) using digital and conventional FURS for kidney stones.
Methods: From September 2007 to April 2011, 118 patients underwent FURS (by the same surgeon). The outcomes were compared between equal numbers of procedures (59 each) using a conventional flexible ureterorenoscope (C-FURS; Olympus URF-P5) and a digital flexible ureterorenoscope (D-FURS; Olympus URF-V). Although the deflection, working channel, and field view are similar in both, the initial and terminal diameter is 8.4F and 9.9F and 6.9F and 8.4F for the D-FURS and C-FURS, respectively. The mean stone fragmentation time was calculated by the size per operative time. The preoperative, operative, and postoperative data were retrospectively analyzed and compared.
Results: The patient demographics were comparable. The mean stone size was 12.8 and 12 mm in the C-FURS and D-FURS groups, respectively. The initial assessment of the entire pyelocaliceal system was possible in 58 of 59 cases (98%) in the C-FURS group and 56 of 59 cases (94%) in the D-FURS group. The mean operative time was significantly longer in the C-FURS group (53.8 ± 15.2 minutes vs 44.5 ± 14.9 minutes). The overall stone-free rate 1 month after the procedure was 86% in the C-FURS group and 88% in the D-FURS group.
Conclusion: Although on comparison, the D-FURS had slightly limited maneuverability, comparable success rates can be achieved with both conventional and digital ureteroscopes. D-FURSs significantly reduced the operative time compared with C-FURSs.