Comparison of fatigue strength of C2 pedicle screws, C2 pars screws, and a hybrid construct in C1-C2 fixation

Spine (Phila Pa 1976). 2014 Jan 1;39(1):E12-9. doi: 10.1097/BRS.0000000000000063.

Abstract

Study design: A biomechanical study comparing the fatigue strength of different types of C2 fixation in a C1-C2 construct.

Objective: To determine the pullout strength of a C2 pedicle screw and C2 pars screw after cyclical testing and differentiate differences in stiffness pre- and post-cyclical loading of 3 different C1-C2 fixations.

Summary of background data: Some surgeons use a short C2 pars screw in a C1-C2 construct, because it is less technically demanding and/or when the vertebral artery is high riding. Difference in construct stiffness between use of bilateral C2 pedicle screws, bilateral C2 pars screws, or a hybrid construct is unknown.

Methods: Biomechanical testing was performed on 15 specimens. A bicortical C1 lateral mass screw was used in combination with 1 of 3 methods of C2 fixation: (1) bilateral long C2 pedicle screws (LL), (2) bilateral 14-mm C2 pars screws (SS), and (3) unilateral long C2 pedicle screw with a contralateral 14-mm C2 pars screw (LS). Each construct was subject to 16,000 cycles to simulate the immediate postoperative period. Changes in motion in flexion-extension, lateral bending, and axial rotation were calculated. This was followed by pullout testing.

Results: The ability to limit range of motion significantly decreased after cyclical testing in flexion-extension, lateral bending, and axial rotation for all 3 groups. After loading, the LL and LS groups had less percentage of increase in motion in flexion-extension and lateral bending than the SS group. Overall, the average pullout strength of a pedicle screw was 92% stronger than a pars screw.

Conclusion: C2 pedicle screws have twice the pullout strength of C2 pars screws after cyclical loading. In cases in which the anatomy limits placement of bilateral C2 pedicle screws, a construct using a unilateral C2 pedicle screw with a contralateral short pars screw is a viable option and compares favorably with a bilateral C2 pedicle screw construct.

Level of evidence: N/A.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Atlanto-Axial Joint / surgery*
  • Axis, Cervical Vertebra / surgery*
  • Biomechanical Phenomena
  • Bone Screws
  • Cervical Atlas / surgery*
  • Female
  • Humans
  • Joint Instability / surgery*
  • Male
  • Middle Aged
  • Range of Motion, Articular
  • Spinal Fusion / methods*