Laser interstitial thermotherapy (LITT) is a minimally invasive technique used to thermally destroy tumour cells. Being based on hyperthermia, LITT outcome depends on the temperature distribution inside the tissue. Recently, CT scan thermometry, based on the dependence of the CT number (HU) on tissue temperature (T) has been introduced during LITT; it is an attractive approach to monitor T because it overcomes the concerns related to the invasiveness. We performed LITT on nine ex vivo swine livers at three different laser powers, (P=1.5 W, P=3 W, P=5 W) with a constant treatment time t=200 s; HU is averaged on two ellipsoidal regions of interest (ROI) of 0.2 cm2, placed at two distances from the applicator (d=3.6 mm and d=8.7 mm); a reference ROI was placed away from the applicator (d=30 mm). The aim of this study is twofold: 1) to evaluate the effect of the T increase in terms of HU variation in ex vivo swine livers undergoing LITT; and 2) to estimate the P value for tissue vaporization. To the best of our knowledge, this is the first study focused on the HU variation in swine livers undergoing LITT at different P. The reported findings could be useful to assess the effect of LITT on the liver in terms of both T changes and tissue vaporization, with the aim to obtain an effective therapy.