Evaluation of manual and automatic manually triggered ventilation performance and ergonomics using a simulation model

Respir Care. 2014 May;59(5):735-42. doi: 10.4187/respcare.02557. Epub 2013 Oct 29.

Abstract

Background: In the absence of endotracheal intubation, the manual bag-valve-mask (BVM) is the most frequently used ventilation technique during resuscitation. The efficiency of other devices has been poorly studied. The bench-test study described here was designed to evaluate the effectiveness of an automatic, manually triggered system, and to compare it with manual BVM ventilation.

Methods: A respiratory system bench model was assembled using a lung simulator connected to a manikin to simulate a patient with unprotected airways. Fifty health-care providers from different professional groups (emergency physicians, residents, advanced paramedics, nurses, and paramedics; n = 10 per group) evaluated manual BVM ventilation, and compared it with an automatic manually triggered device (EasyCPR). Three pathological situations were simulated (restrictive, obstructive, normal). Standard ventilation parameters were recorded; the ergonomics of the system were assessed by the health-care professionals using a standard numerical scale once the recordings were completed.

Results: The tidal volume fell within the standard range (400-600 mL) for 25.6% of breaths (0.6-45 breaths) using manual BVM ventilation, and for 28.6% of breaths (0.3-80 breaths) using the automatic manually triggered device (EasyCPR) (P < .0002). Peak inspiratory airway pressure was lower using the automatic manually triggered device (EasyCPR) (10.6 ± 5 vs 15.9 ± 10 cm H2O, P < .001). The ventilation rate fell consistently within the guidelines, in the case of the automatic manually triggered device (EasyCPR) only (10.3 ± 2 vs 17.6 ± 6, P < .001). Significant pulmonary overdistention was observed when using the manual BVM device during the normal and obstructive sequences. The nurses and paramedics considered the ergonomics of the automatic manually triggered device (EasyCPR) to be better than those of the manual device.

Conclusions: The use of an automatic manually triggered device may improve ventilation efficiency and decrease the risk of pulmonary overdistention, while decreasing the ventilation rate.

Keywords: automated system; bag-valve-mask; ergonomy; manual ventilation; performance evaluation; simulation.

Publication types

  • Evaluation Study

MeSH terms

  • Attitude of Health Personnel*
  • Cardiopulmonary Resuscitation / adverse effects
  • Cardiopulmonary Resuscitation / instrumentation*
  • Cardiopulmonary Resuscitation / methods
  • Equipment Design
  • Ergonomics
  • Humans
  • Intermittent Positive-Pressure Breathing / adverse effects
  • Intermittent Positive-Pressure Breathing / instrumentation*
  • Intermittent Positive-Pressure Breathing / methods
  • Manikins
  • Residual Volume
  • Respiratory Rate
  • Tidal Volume