Accuracy of postcontrast 3D turbo spin-echo MR sequence for the detection of enhanced inflammatory lesions in patients with multiple sclerosis

AJNR Am J Neuroradiol. 2014 Mar;35(3):519-23. doi: 10.3174/ajnr.A3795. Epub 2013 Nov 7.

Abstract

Background and purpose: Therapeutic strategies for patients with MS partly rely on contrast-enhanced MR imaging. Our aim was to assess the diagnostic performance of 3D turbo spin-echo MR imaging with variable refocusing flip angles at 3T for the detection of enhanced inflammatory lesions in patients with multiple sclerosis.

Materials and methods: Fifty-six patients with MS were prospectively investigated by using postcontrast T1-weighted axial 2D spin-echo and 3D TSE MR images. The order in which both sequences were performed was randomized. Axial reformats from 3D T1 TSE were generated to match the 2D spin-echo images. The reference standard was defined by using clinical data and all MR images available. Three separate sets of MR images (2D spin-echo images, axial reformats, and multiplanar images from 3D TSE sequences) were examined in a blinded fashion by 2 neuroradiologists separately for the detection of enhanced MS lesions. Image artifacts and contrast were evaluated.

Results: No artifacts related to vascular pulsation were observed on 3D TSE images, whereas image artifacts were demonstrated on 2D spin-echo images in 41 patients. One hundred twelve enhanced MS lesions were identified in 19 patients. Sixty-four lesions were correctly diagnosed by using 2D spin-echo images; 90, by using 3D TSE axial reformatted views; and 106, by using multiplanar analysis of the 3D TSE sequence. Multiplanar analysis was 94.7% sensitive and 100% specific for the diagnosis of patients with at least 1 enhanced lesion. Contrast of enhanced MS lesions was significantly improved by using the 3D TSE sequence (P < .011).

Conclusions: The 3D TSE sequence with multiplanar analysis is a useful tool for the detection of enhanced MS lesions.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Dimensional Measurement Accuracy
  • Encephalitis / diagnosis*
  • Encephalitis / etiology
  • Female
  • Humans
  • Image Enhancement
  • Imaging, Three-Dimensional*
  • Magnetic Resonance Imaging* / methods
  • Male
  • Multiple Sclerosis / complications
  • Multiple Sclerosis / diagnosis*
  • Neuroimaging / methods*
  • Prospective Studies