Compensation of signal loss due to cardiac motion in point-resolved spectroscopy of the heart

Magn Reson Med. 2014 Nov;72(5):1201-7. doi: 10.1002/mrm.25028. Epub 2013 Nov 20.

Abstract

Purpose: To study the influence of cardiac motion on point-resolved single voxel spectroscopy (PRESS) of the human heart and to propose an improved sequence design to compensate for cardiac motion-related signal losses.

Methods: Numerical simulations and in vivo measurements were performed to study motion effects in cardiac spectroscopy. Based on the findings, an improved PRESS sequence with reduced free induction decay spoiling areas was implemented. Myocardial triglyceride (TG) content as estimated with the improved PRESS sequence was compared with the default PRESS sequence commonly used for spectroscopy in nonmoving tissues that was provided by the manufacturer.

Results: Numerical simulations and in vivo data show the importance of optimal trigger delays to avoid signal losses in cardiac spectroscopy. The improved PRESS sequence was found to yield a significantly reduced sensitivity to cardiac motion and trigger delay setting and provided a 1.7-fold increase in signal-to-noise ratio on average compared with the default PRESS sequence. Intrasubject variation of myocardial TG content was reduced with the improved sequence compared with the default PRESS sequence (coefficient of variation: 0.11 ± 0.03 versus 0.23 ± 0.06).

Conclusion: Improvement of the sequence design is essential to reduce the sensitivity of PRESS to cardiac motion.

Keywords: cardiac motion; cardiac spectroscopy; creatine; motion compensation; signal-to-noise ratio; triglyceride.

Publication types

  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Adult
  • Computer Simulation
  • Female
  • Healthy Volunteers
  • Heart / physiology*
  • Humans
  • Magnetic Resonance Spectroscopy / methods*
  • Male
  • Movement
  • Signal-To-Noise Ratio
  • Triglycerides / metabolism

Substances

  • Triglycerides