Development of a prototype PET scanner with depth-of-interaction measurement using solid-state photomultiplier arrays and parallel readout electronics

Phys Med Biol. 2014 Mar 7;59(5):1223-38. doi: 10.1088/0031-9155/59/5/1223. Epub 2014 Feb 20.

Abstract

In this study, we developed a prototype animal PET by applying several novel technologies to use solid-state photomultiplier (SSPM) arrays to measure the depth of interaction (DOI) and improve imaging performance. Each PET detector has an 8 × 8 array of about 1.9 × 1.9 × 30.0 mm(3) lutetium-yttrium-oxyorthosilicate scintillators, with each end optically connected to an SSPM array (16 channels in a 4 × 4 matrix) through a light guide to enable continuous DOI measurement. Each SSPM has an active area of about 3 × 3 mm(2), and its output is read by a custom-developed application-specific integrated circuit to directly convert analogue signals to digital timing pulses that encode the interaction information. These pulses are transferred to and are decoded by a field-programmable gate array-based time-to-digital convertor for coincident event selection and data acquisition. The independent readout of each SSPM and the parallel signal process can significantly improve the signal-to-noise ratio and enable the use of flexible algorithms for different data processes. The prototype PET consists of two rotating detector panels on a portable gantry with four detectors in each panel to provide 16 mm axial and variable transaxial field-of-view (FOV) sizes. List-mode ordered subset expectation maximization image reconstruction was implemented. The measured mean energy, coincidence timing and DOI resolution for a crystal were about 17.6%, 2.8 ns and 5.6 mm, respectively. The measured transaxial resolutions at the center of the FOV were 2.0 mm and 2.3 mm for images reconstructed with and without DOI, respectively. In addition, the resolutions across the FOV with DOI were substantially better than those without DOI. The quality of PET images of both a hot-rod phantom and mouse acquired with DOI was much higher than that of images obtained without DOI. This study demonstrates that SSPM arrays and advanced readout/processing electronics can be used to develop a practical DOI-measureable PET scanner.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Equipment Design
  • Equipment Failure Analysis
  • Imaging, Three-Dimensional / instrumentation*
  • Imaging, Three-Dimensional / veterinary*
  • Mice
  • Mice, Nude
  • Photometry / instrumentation*
  • Photometry / veterinary*
  • Positron-Emission Tomography / instrumentation*
  • Positron-Emission Tomography / veterinary*
  • Reproducibility of Results
  • Semiconductors
  • Sensitivity and Specificity
  • Signal Processing, Computer-Assisted / instrumentation
  • User-Computer Interface*