Aim: To assess the value of dual-energy computed tomography (DECT) and an iterative frequency split-normalized metal artefact reduction (IFS-MAR) algorithm compared to filtered back projections (FBP) from single-energy CT (SECT) for artefact reduction in internally fixated humeral fractures.
Materials and methods: Six internally fixated cadaveric humeri were examined using SECT and DECT. Data were reconstructed using FBP, IFS-MAR, and mono-energetic DECT extrapolations. Image analysis included radiodensity values and qualitative evaluation of artefacts, image quality, and level of confidence for localizing screw tips.
Results: Radiodensity values of streak artefacts were significantly different (p < 0.05) between FBP (-104 ± 222) and IFS-MAR (73 ± 122), and between FBP and DECT (32 ± 151), without differences between IFS-MAR and DECT (p < 0.553). Compared to FBP, qualitative artefacts were significantly reduced using IFS-MAR (p < 0.001) and DECT (p < 0.05), without significant differences between IFS-MAR and DECT (p < 0.219). Image quality significantly (p = 0.016) improved for IFS-MAR and DECT compared to FBP, without significant differences between IFS-MAR and DECT (p < 0.553). The level of confidence for screw tip localization was assessed as best for DECT in all cases.
Conclusion: Both IFS-MAR in SECT and mono-energetic DECT produce improved image quality and a reduction of metal artefacts. Screw tip positions can be most confidently assessed using DECT.
Copyright © 2014 The Royal College of Radiologists. Published by Elsevier Ltd. All rights reserved.