Percutaneous mitral valve repair with the MitraClip system according to the predicted risk by the logistic EuroSCORE: preliminary results from the German Transcatheter Mitral Valve Interventions (TRAMI) Registry

Catheter Cardiovasc Interv. 2014 Oct 1;84(4):591-8. doi: 10.1002/ccd.25493. Epub 2014 Apr 30.

Abstract

Objective: To evaluate in-hospital and short-term outcomes of percutaneous mitral valve repair according to patients' logistic EuroSCORE (logEuroSCORE) in a multicenter registry

Background: The logEuroSCORE is an established tool to predict the risk of mortality during cardiac surgery. In high-risk patients percutaneous mitral valve repair with the MitraClip system represents a less-invasive alternative

Methods: Data from 1002 patients, who underwent percutaneous mitral valve repair with the MitraClip system, were analyzed in the German Transcatheter Mitral Valve Interventions (TRAMI) Registry. A logEuroSCORE (mortality risk in %) ≥ 20 was considered high risk

Results: Of all patients, 557 (55.6%) had a logEuroSCORE ≥ 20. Implantation of the MitraClip was successful in 95.5 % (942/986) patients. Moderate residual mitral valve regurgitation was more often detected in patients with a logEuroSCORE ≥ 20 (23.8% vs. 17.1%, respectively, P < 0.05). In patients with a logEuroSCORE ≥ 20 the procedural complication rate was 8.9% (vs. 6.4, n.s.) and the in-hospital MACCE rate 4.9% (vs. 1.4% P < 0.01). The in-hospital mortality rate in patients with a logEuroSCORE ≥ 20 and logEuroSCORE < 20 was 4.3 and 1.1%, respectively (P ≤ 0.01) CONCLUSION: Percutaneous mitral valve repair with the MitraClip system is feasible in patients with a logEuroSCORE ≥ 20 with similar procedural results compared to patients with lower predicted risk. Although mortality was four times higher than in patients with logEuroSCORE < 20, mortality in high risk patients was lower than predicted. In those with a logEuroSCORE ≥ 20, moderate residual mitral valve regurgitation was more frequent.

Keywords: MitraClip; inoperable; logistic EuroScore; mitral valve regurgitation; percutaneous mitral valve repair.

Publication types

  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Cardiac Catheterization / adverse effects
  • Cardiac Catheterization / instrumentation*
  • Cardiac Catheterization / mortality
  • Chi-Square Distribution
  • Feasibility Studies
  • Female
  • Germany
  • Heart Valve Prosthesis Implantation / adverse effects
  • Heart Valve Prosthesis Implantation / instrumentation*
  • Heart Valve Prosthesis Implantation / mortality
  • Heart Valve Prosthesis*
  • Hospital Mortality
  • Humans
  • Logistic Models
  • Male
  • Mitral Valve Insufficiency / diagnosis
  • Mitral Valve Insufficiency / mortality
  • Mitral Valve Insufficiency / physiopathology
  • Mitral Valve Insufficiency / therapy*
  • Mitral Valve* / physiopathology
  • Patient Selection
  • Prospective Studies
  • Prosthesis Design
  • Registries
  • Retrospective Studies
  • Risk Assessment
  • Risk Factors
  • Time Factors
  • Treatment Outcome