Background: Fully covered esophageal self-expandable metal stents (SEMSs) are potentially removable but can be associated with high migration rates. For precise positioning, non-foreshortening SEMSs are preferred. Recently, a new fully covered non-foreshortening SEMS with anti-migration features was introduced.
Objective: To evaluate the efficacy and safety of this new esophageal SEMS.
Design: Retrospective study.
Setting: Single, tertiary-care center.
Patients: Consecutive patients with malignant and benign strictures with dysphagia grade of ≥3 and patients with fistulas/leaks were studied.
Interventions: Stent placement and removal.
Main outcome measurements: Technical success in stent deployment/removal, efficacy in relieving dysphagia and sealing fistulas/leaks, and adverse events.
Results: Forty-three stents were placed in 35 patients (mean [± standard deviation] age 65 ± 11 years; 31 male), 24 for malignant and 11 for benign (5 strictures, 6 leaks) indications. Technical success in precise SEMS placement was 100%. The after-stent dysphagia grade improved significantly (at 1 week: 1.5 ± 0.7; at 4 weeks: 1.2 ± 0.4; baseline: 3.8 ± 0.4; P < .0001). Twenty stents were removed for clinical indications, with technical success of 100%. All leaks sealed after SEMS placement and did not recur after stent removal. All benign strictures recurred after stent removal. Adverse events included migration (14%), chest pain (11%), and dysphagia from tissue hyperplasia (6%). There was no stent-related mortality.
Limitations: Nonrandomized, single-center study.
Conclusion: The new esophageal SEMS was effective in relieving malignant dysphagia, allowed for precise placement, and was easily removable. It was effective in treating benign esophageal fistulas and leaks. Stent-related adverse events were acceptable.
Copyright © 2014 American Society for Gastrointestinal Endoscopy. Published by Elsevier Inc. All rights reserved.