Systematic Comparison of the Performance of Integrated Whole-Body PET/MR Imaging to Conventional PET/CT for ¹⁸F-FDG Brain Imaging in Patients Examined for Suspected Dementia

J Nucl Med. 2014 Jun;55(6):923-31. doi: 10.2967/jnumed.113.126813. Epub 2014 May 15.

Abstract

Technologic specifications of recently introduced integrated PET/MR instrumentation, such as MR-based attenuation correction, may particularly affect brain imaging procedures. To evaluate the qualitative performance of PET/MR in clinical neuroimaging, we systematically compared results obtained with integrated PET/MR with conventional PET/CT in the same patients examined for assessment of cognitive impairment.

Methods: Thirty patients underwent a single-injection ((18)F-FDG), dual-imaging protocol including PET/CT and integrated PET/MR imaging in randomized order. Attenuation and scatter correction were performed using low-dose CT for the PET/CT and segmented Dixon MR imaging data for the PET/MR. Differences between PET/MR and PET/CT were assessed via region-of-interest (ROI)-based and voxel-based statistical group comparison. Analyses involved attenuation-corrected (AC) and non-attenuation-corrected (NAC) data. Individual PET/MR and PET/CT datasets were compared versus a predefined independent control population, using 3-dimensional stereotactic surface projections.

Results: Generally, lower measured PET signal values were obtained throughout the brain in ROI-based quantification of the PET signal for PET/MR as compared with PET/CT in AC and NAC data, independently of the scan order. After elimination of global effects, voxel-based and ROI-based group comparison still revealed significantly lower relative tracer signal in PET/MR images in frontoparietal portions of the neocortex but significantly higher relative signal in subcortical and basal regions of the brain than the corresponding PET/CT images of the AC data. In the corresponding NAC images, the discrepancies in frontoparietal portions of the neocortex were diminished, but the subcortical overestimation of tracer intensity by PET/MR persisted.

Conclusion: Considerable region-dependent differences were observed between brain imaging data acquired on the PET/MR, compared with corresponding PET/CT images, in patients evaluated for neurodegenerative disorders. These findings may only in part be explained by inconsistencies in the attenuation-correction procedures. The observed differences may interfere with semiquantitative evaluation and with individual qualitative clinical assessment and they need to be considered, for example, for clinical trials. Improved attenuation-correction algorithms and a PET/MR-specific healthy control database are recommended for reliable and consistent application of PET/MR for clinical neuroimaging.

Keywords: Dixon MRI sequence; PET/CT; PET/MR; brain imaging; neurodegeneration.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Brain / diagnostic imaging*
  • Dementia / diagnosis*
  • Dementia / diagnostic imaging
  • Female
  • Fluorodeoxyglucose F18*
  • Humans
  • Image Processing, Computer-Assisted
  • Magnetic Resonance Imaging
  • Male
  • Middle Aged
  • Multimodal Imaging / methods*
  • Phantoms, Imaging
  • Positron-Emission Tomography
  • Tomography, X-Ray Computed
  • Whole Body Imaging*

Substances

  • Fluorodeoxyglucose F18