Iterative model reconstruction: improved image quality of low-tube-voltage prospective ECG-gated coronary CT angiography images at 256-slice CT

Eur J Radiol. 2014 Aug;83(8):1408-15. doi: 10.1016/j.ejrad.2014.04.027. Epub 2014 May 2.

Abstract

Objectives: To investigate the effects of a new model-based type of iterative reconstruction (M-IR) technique, the iterative model reconstruction, on image quality of prospectively gated coronary CT angiography (CTA) acquired at low-tube-voltage.

Methods: Thirty patients (16 men, 14 women; mean age 52.2 ± 13.2 years) underwent coronary CTA at 100-kVp on a 256-slice CT. Paired image sets were created using 3 types of reconstruction, i.e. filtered back projection (FBP), a hybrid type of iterative reconstruction (H-IR), and M-IR. Quantitative parameters including CT-attenuation, image noise, and contrast-to-noise ratio (CNR) were measured. The visual image quality, i.e. graininess, beam-hardening, vessel sharpness, and overall image quality, was scored on a 5-point scale. Lastly, coronary artery segments were evaluated using a 4-point scale to investigate the assessability of each segment.

Results: There was no significant difference in coronary arterial CT attenuation among the 3 reconstruction methods. The mean image noise of FBP, H-IR, and M-IR images was 29.3 ± 9.6, 19.3 ± 6.9, and 12.9 ± 3.3 HU, respectively, there were significant differences for all comparison combinations among the 3 methods (p<0.01). The CNR of M-IR was significantly better than of FBP and H-IR images (13.5 ± 5.0 [FBP], 20.9 ± 8.9 [H-IR] and 39.3 ± 13.9 [M-IR]; p<0.01). The visual scores were significantly higher for M-IR than the other images (p<0.01), and 95.3% of the coronary segments imaged with M-IR were of assessable quality compared with 76.7% of FBP- and 86.9% of H-IR images.

Conclusions: M-IR can provide significantly improved qualitative and quantitative image quality in prospectively gated coronary CTA using a low-tube-voltage.

Keywords: Coronary CT angiography; Image quality; Iterative reconstruction; Low tube voltage; Radiation dose.

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Cardiac-Gated Imaging Techniques / methods*
  • Contrast Media
  • Coronary Angiography / methods*
  • Female
  • Humans
  • Iohexol
  • Male
  • Middle Aged
  • Quality Improvement
  • Radiographic Image Interpretation, Computer-Assisted / methods*
  • Retrospective Studies
  • Tomography, X-Ray Computed / methods*

Substances

  • Contrast Media
  • Iohexol