Background: Postoperative lead perforation is a life-threatening complication of cardiac pacing. Identification of precipitating factors for this serious complication may help to anticipate a specific risk profile and to minimize the incidence.
Methods: We conducted a retrospective tertiary referral center analysis to clarify clinical, anatomical, and technical characteristics related to pacemaker (PM) and cardioverter/defibrillator lead perforation. We examined the baseline characteristics and the symptoms. In a subgroup, we investigated the myocardial thickness on contrast-enhanced cardiac computed tomography.
Results: We enrolled 26 patients. Female gender appears to put patients at slightly increased risk for lead perforation. In a majority active fixation leads were used. Symptoms occurred in 72%. Pericardial effusion and tamponade were present in 38% and 19%, respectively. Sensing was compromised in 65%. A high pacing threshold or exit block occurred in 92%. Myocardial thickness did not differ between patients with or without perforation. In 96%, the perforation was treated by transvenous withdrawal.
Conclusion: Chest pain, phrenic stimulation, bad sensing, or exit block early after PM implantation must prompt radiological and echocardiographic evaluation. A missing pericardial effusion particularly late after implantation does not rule out a perforation. Especially active fixating leads have a higher risk of perforation. With cardiac surgery in standby transvenous withdrawal is a safe way to treat lead perforation.
Keywords: cardioverter/defibrillator; lead perforation; pacemaker; risk factors.
©2014 Wiley Periodicals, Inc.