Background: Nobori is a novel biolimus A9-eluting stent (BES) coated with a biodegradable polymer only on the abluminal side, which degrades over 6-9 months post-stent deployment. The course of vessel reaction after deployment at these time points remains unclear.
Methods and results: We serially evaluated 28 BES implanted in de novo coronary lesions of 23 patients using optical coherence tomography (OCT) at 6 and 12 months post-stenting. Standard OCT variables, the percentage of stent with peri-strut low-intensity area (PLIA, a region around stent struts homogenously showing lesser intensity than the surrounding tissue, suggesting fibrin deposition or impaired neointima maturation) and that with in-stent thrombi were evaluated. There was a significant, but small increase in neointimal thickness (72 ± 23 to 82 ± 25 µm, P=0.006) from the 6- to the 12-month follow-up, without a significant decrease in minimum lumen area (P=0.30). The incidences of uncovered and malapposed struts were low at 6 months and reduced further at 12 months (3.96 ± 3.97% to 1.51 ± 1.63%, P=0.001, and 0.50 ± 1.84% to 0.06 ± 0.24%, P=0.20, respectively). The frequency of stent with PLIA decreased during the follow-up (57% to 32%, P=0.05) and that with in-stent thrombi also numerically decreased (7% to 0%, P=0.24).
Conclusions: Neointimal hyperplasia was persistently suppressed following BES implantation up to 12 months. Simultaneously, favorable vessel healing was achieved at 6 months without a delaying adverse reaction for up to 12 months.