Methods: 17 physicians, experienced in transvenous lead removal, performed a lead extraction manoeuvre of an ICD lead on a torso phantom. They were advised to stop traction only when further traction would be considered as harmful to the patient or when--based on their experience--a change in the extraction strategy was indicated. Traction forces were recorded with a digital precision gauge.
Results: Median traction forces on the endocardium were 10.9 N (range from 3.0 N to 24.7 N and interquartile range from 7.9 to 15.3). Forces applied to the proximal end were estimated to be 10% higher than those measured at the tip of the lead due to a friction loss.
Conclusion: A traction force of around 11 N is typically exerted during standard transvenous extraction of ICD leads. A traction threshold for a safe procedure derived from a pool of experienced extractionists may be helpful for the development of required adequate simulator trainings.