Abstract

Determination of cholic acid is very important and necessary in body fluids due to its both pharmaceutical and clinical significance. In this study, a quartz crystal microbalance (QCM) nanosensor, which is imprinted cholic acid, has been developed for the assignation of cholic acid. The cholic acid selective memories have been generated on QCM electrode surface by using molecularly imprinted polymer (MIP) based on methacryloylamidohistidine-copper (II) (MAH-Cu(II)) pre-organized monomer. The cholic acid imprinted nanosensor was characterized by atomic force microscopy (AFM) and then analytical performance of the cholic acid imprinted QCM nanosensor was studied. The detection limit was found to be 0.0065μM with linear range of 0.01-1,000 μM. Moreover, the high value of Langmuir constant (b) (7.3*10(5)) obtained by Langmuir graph showed that the cholic acid imprinted nanosensor had quite strong binding sites affinity. At the last step of this procedure, cholic acid levels in body fluids were determined by the prepared imprinted QCM nanosensor.

Keywords: Cholic acid; MIP; Nanosensor; QCM.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Cholic Acid / blood*
  • Cholic Acid / chemistry
  • Cholic Acid / urine*
  • Humans
  • Limit of Detection
  • Microscopy, Atomic Force
  • Molecular Imprinting / instrumentation
  • Molecular Imprinting / methods*
  • Nanotechnology / instrumentation
  • Nanotechnology / methods*
  • Quartz Crystal Microbalance Techniques / instrumentation
  • Quartz Crystal Microbalance Techniques / methods*
  • Surface Properties

Substances

  • Cholic Acid