Risk factors for erosion of artificial urinary sphincters: a multicenter prospective study

Urology. 2014 Oct;84(4):934-8. doi: 10.1016/j.urology.2014.05.043. Epub 2014 Aug 8.

Abstract

Objective: To evaluate the short- to medium-term outcomes after artificial urinary sphincter (AUS) placement from a large, multi-institutional, prospective, follow-up study. We hypothesize that along with radiation, patients with any history of a direct surgery to the urethra will have higher rates of eventual AUS explantation for erosion and/or infection.

Materials and methods: A prospective outcome analysis was performed on 386 patients treated with AUS placement from April 2009 to December 2012 at 8 institutions with at least 3 months of follow-up. Charts were analyzed for preoperative risk factors and postoperative complications requiring explantation.

Results: Approximately 50% of patients were considered high risk. High risk was defined as patients having undergone radiation therapy, urethroplasty, multiple treatments for bladder neck contracture or urethral stricture, urethral stent placement, or a history of erosion or infection in a previous AUS. A total of 31 explantations (8.03%) were performed during the follow-up period. Overall explantation rates were higher in those with prior radiation and prior UroLume. Men with prior AUS infection or erosion also had a trend for higher rates of subsequent explantation. Men receiving 3.5-cm cuffs had significantly higher explantation rates than those receiving larger cuffs.

Conclusion: This outcomes study confirms that urethral risk factors, including radiation history, prior AUS erosion, and a history of urethral stent placement, increase the risk of AUS explantation in short-term follow-up.

Publication types

  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Follow-Up Studies
  • Humans
  • Prospective Studies
  • Prosthesis Failure*
  • Risk Factors
  • Urinary Sphincter, Artificial / adverse effects*