Factors affecting perceptual thresholds in a suprachoroidal retinal prosthesis

Invest Ophthalmol Vis Sci. 2014 Sep 9;55(10):6467-81. doi: 10.1167/iovs.14-14396.

Abstract

Purpose: The suprachoroidal location for a retinal prosthesis provides advantages over other locations in terms of a simplified surgical procedure and a potentially more stable electrode-neural interface. The aim of this study was to assess the factors affecting perceptual thresholds, and to optimize stimulus parameters to achieve the lowest thresholds in patients implanted with a suprachoroidal retinal prosthesis.

Methods: Three patients with profound vision loss from retinitis pigmentosa were implanted with a suprachoroidal array. Perceptual thresholds measured on individual electrodes were analyzed as a function of stimulus (return configuration, pulse polarity, pulse width, interphase gap, and rate), electrode (area and number of ganged electrodes), and clinical (retinal thickness and electrode-retina distance) parameters.

Results: A total of 92.8% of 904 measurements made up to 680 days post implantation yielded thresholds (range, 44-436 nanocoulombs [nC]) below the safe charge limit. Thresholds were found to vary between individuals and to depend significantly on electrode-retina distance, negligibly on retinal thickness, and not on electrode area or the number of ganged electrodes. Lowest thresholds were achieved when using a monopolar return, anodic-first polarity, short pulse widths (100 μs) combined with long interphase gaps (500 μs), and high stimulation rates (≥400 pulses per second [pps]).

Conclusions: With suprachoroidal stimulation, anodic-first pulses with a monopolar return are most efficacious. To enable high rates, an appropriate combination of pulse width and interphase gap must be chosen to ensure low thresholds and electrode voltages. Electrode-retina distance needs to be monitored carefully owing to its influence on thresholds. These results inform implantable stimulator specifications for a suprachoroidal retinal prosthesis. (ClinicalTrials.gov number, NCT01603576.).

Keywords: retinal prosthesis; retinitis pigmentosa; suprachoroidal; threshold.

Publication types

  • Clinical Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Electric Stimulation
  • Evoked Potentials, Visual / physiology*
  • Female
  • Humans
  • Male
  • Microelectrodes
  • Middle Aged
  • Prosthesis Design
  • Retina / physiopathology
  • Retina / surgery*
  • Retinitis Pigmentosa / physiopathology
  • Retinitis Pigmentosa / surgery*
  • Treatment Outcome
  • Visual Cortex / physiopathology*
  • Visual Perception / physiology*
  • Visual Prosthesis*
  • Young Adult

Associated data

  • ClinicalTrials.gov/NCT01603576