A randomized comparison of novel biodegradable polymer- and durable polymer-coated cobalt-chromium sirolimus-eluting stents

JACC Cardiovasc Interv. 2014 Dec;7(12):1352-60. doi: 10.1016/j.jcin.2014.09.001. Epub 2014 Nov 12.

Abstract

Objectives: The aim of this study was to investigate the hypothesis that a novel biodegradable polymer-coated, cobalt-chromium (CoCr), sirolimus-eluting stent (BP-SES) is noninferior in safety and efficacy outcomes compared with a durable polymer (DP)-SES.

Background: No randomized trials have the compared safety and efficacy of BP-SES versus DP-SES on similar CoCr platforms, thereby isolating the effect of the polymer type.

Methods: In this prospective, single-blind, randomized trial conducted at 32 Chinese sites, 2,737 patients eligible for coronary stenting were treated with BP- or DP-SES in a 2:1 ratio. The primary endpoint was 12-month target lesion failure (TLF), a composite of cardiac death, target vessel myocardial infarction, or clinically indicated target lesion revascularization. Secondary endpoints included TLF components, and definite/probable stent thrombosis.

Results: At 12 months, the difference in the primary endpoint of TLF between BP-SES (6.3%) and DP-SES (6.1%) groups was 0.25% (95% confidence interval: -1.67% to 2.17%, p for noninferiority = 0.0002), demonstrating noninferiority of BP-SES to DP-SES. Individual TLF components of cardiac death (0.7% vs. 0.6%, p = 0.62), target vessel myocardial infarction (3.6% vs. 4.3%, p = 0.39), and clinically indicated target lesion revascularization (2.6% vs. 2.2%, p = 0.50) were similar, as were low definite/probable stent thrombosis rates (0.4% vs. 0.6%, p = 0.55).

Conclusions: In this large-scale real-world trial, BP-SES was noninferior to DP-SES for 1-year TLF. (Evaluate Safety and Effectiveness of the Tivoli ® DES and the Firebird ® DES for Treatment of Coronary Revascularization; NCT01681381).

Keywords: biodegradable polymer; clinical outcome; drug-eluting stent; durable polymer; noninferiority trial.

Publication types

  • Comparative Study
  • Multicenter Study
  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Cardiovascular Agents / administration & dosage*
  • China
  • Chromium Alloys*
  • Coronary Artery Disease / diagnosis
  • Coronary Artery Disease / mortality
  • Coronary Artery Disease / therapy*
  • Coronary Thrombosis / etiology
  • Drug-Eluting Stents*
  • Female
  • Humans
  • Kaplan-Meier Estimate
  • Male
  • Middle Aged
  • Myocardial Infarction / diagnosis
  • Myocardial Infarction / mortality
  • Myocardial Infarction / therapy*
  • Percutaneous Coronary Intervention / adverse effects
  • Percutaneous Coronary Intervention / instrumentation*
  • Percutaneous Coronary Intervention / mortality
  • Polymers*
  • Proportional Hazards Models
  • Prospective Studies
  • Prosthesis Design
  • Recurrence
  • Risk Factors
  • Single-Blind Method
  • Sirolimus / administration & dosage*
  • Time Factors
  • Treatment Outcome

Substances

  • Cardiovascular Agents
  • Chromium Alloys
  • Polymers
  • Sirolimus

Associated data