Background: The aim of this experimental study was to assess the feasibility of complete endovascular arch reconstruction by in situ retrograde fenestration and to investigate the impact of stent-graft material on stent-graft fenestrations.
Methods: The experiments were performed using 8 cadaveric human thoracic aortas (aortic arch) using 2 different stent-graft types: woven polyester (Valiant Captivia; Medtronic Vascular, Santa Rosa, CA) and expanded polytetrafluoroethylene (conformable [C]-TAG; W.L. Gore & Associates, Flagstaff, AZ). A benchtop aortic pulsatile flow model was used. Stent-grafts were deployed into the aortic arch, covering the ostia of the supraaortic trunks. A 5-mm 30-degree angioscope was introduced into the ascending aorta to monitor the procedure. Retrograde fenestration and deployment of the balloon expandable stent-graft was performed sequentially for each supraaortic trunk. Subsequent to stent-graft explantation, macroscopic evaluation of each fenestration was performed.
Results: All attempts to fenestrate the C-TAG and Valiant stent-grafts and implant the covered stent through the supraaortic trunks were successful. In all cases, branch stents were patent and no endoleak was evident. The Valiant stent-graft was easier to puncture because of the higher radial force of the stent-graft providing better counterpressure; however, stent-graft material had no impact on the quality of fenestrations.
Conclusions: Total endovascular repair of the aortic arch through in situ retrograde fenestration of stent-grafts is feasible. The behavior of the 2 types of stent-graft was significantly different while the fenestrations were fashioned, but stent-graft material had no impact on the quality of fenestrations.
Copyright © 2014 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.