Evaluation of a chest circumference-adapted protocol for low-dose 128-slice coronary CT angiography with prospective electrocardiogram triggering

Korean J Radiol. 2015 Jan-Feb;16(1):13-20. doi: 10.3348/kjr.2015.16.1.13. Epub 2015 Jan 9.

Abstract

Objective: To assess the effect of chest circumference-adapted scanning protocol on radiation exposure and image quality in patients undergoing prospective electrocardiogram (ECG)-triggered coronary CT angiography (CCTA).

Materials and methods: One hundred-eighty-five consecutive patients, who had undergone prospective ECG triggering CCTA with a 128-slice CT, were included in the present study. Nipple-level chest circumference, body weight and height were measured before CT examinations. Patients were divided into four groups based on kV/ref · mAs = 100/200, 100/250, 120/200, and 120/250, when patient's chest circumference was ≤ 85.0 (n = 56), 85.0-90.0 (n = 53), 90.0-95.0 (n = 44), and > 95.0 (n = 32), respectively. Image quality per-segment was independently assessed by two experienced observers. Image noise and attenuation were also measured. Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated. The effective radiation dose was calculated using CT dose volume index and the dose-length product.

Results: A significant correlation was observed between patients' chest circumference and body mass index (r = 0.762, p < 0.001). Chest circumference ranged from 74 to 105 cm, and the mean effective radiation dose was 1.9-3.8 mSv. Diagnostic image quality was obtained in 98.5% (2440/2478) of all evaluated coronary segments without any significant differences among the four groups (p = 0.650). No significant difference in image noise was observed among the four groups (p = 0.439), thus supporting the validity of the chest circumference-adapted scanning protocol. However, vessel attenuation, SNR and CNR were significantly higher in the 100 kV groups than in the 120 kV groups (p < 0.05).

Conclusion: A measure of chest circumference can be used to adapt tube voltage and current for individualized radiation dose control, with resultant similar image noise and sustained diagnostic image quality.

Keywords: Chest circumference; Computed tomography; Coronary angiography; Image quality; Radiation dose.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Body Mass Index
  • Contrast Media
  • Electrocardiography
  • Female
  • Heart / diagnostic imaging*
  • Heart Rate
  • Humans
  • Male
  • Middle Aged
  • Prospective Studies
  • Radiation Dosage
  • Signal-To-Noise Ratio
  • Tomography, X-Ray Computed / methods*
  • Waist Circumference*

Substances

  • Contrast Media