Three-dimensional ultrasound evaluation of small asymptomatic abdominal aortic aneurysms

Eur J Vasc Endovasc Surg. 2015 Mar;49(3):289-96. doi: 10.1016/j.ejvs.2014.12.022. Epub 2015 Feb 4.

Abstract

Objective: Non-invasive and reproducible size measurements that correlate well with computed tomography (CT) are desirable in the management of small abdominal aortic aneurysms (AAA). Three dimensional ultrasound (3D-US) technology may reduce inaccuracy because of variations in orientation of the image planes and axis. This study aimed to determine any differences in paired size estimation associated with three 3D-US derived methods using 3D-CT as the gold standard. When CTA was not available, the patients were enrolled anyway to assess 3D-US reproducibility in terms of agreement between two physicians.

Methods: In the period from 1 March 2013 to 27 February 2014, consecutive patients with a small AAA, <5.5 cm for men and <5.2 cm for women, underwent 3D-US examination and three AAA size measures were obtained: dual plane diameter, diameter perpendicular to the residual sac's centreline and a partial volume.

Result: In all, 122 consecutive US examinations were performed. Patients were excluded because of inadequate AAA size (n = 11) and for technical reasons (n = 11). Thus, 100 patients (F/M; 20/80) with a median maximum AAA diameter of 46 (range 31-55) mm were analysed. The mean US dual plane diameter and the 3D-US centreline diameter were 2.6 mm and 1.8 mm smaller than the mean 3D-CT centreline diameter, respectively (p = .003). The inter-observer reproducibility coefficient was 3.7 mm for the US dual plane diameter and 3.2 mm for the 3D-US centreline diameter (p = 0.222). For the partial volume, the reproducibility was 8-12%, corresponding to a diameter variability of ±3 mm. The median time used for post-processing of the 3D-US acquisition was 72 (range 46-108) seconds per examination.

Conclusion: 3D-US demonstrated an acceptable reproducibility and a good agreement with 3D-CT, and has the potential to improve future AAA management through more reliable ultrasound guided size estimates.

Keywords: 3D computed tomography; 3D ultrasound; Abdominal; Aortic aneurysm; Three dimensional; Ultrasonography; Volume measurement.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aorta, Abdominal / diagnostic imaging*
  • Aortic Aneurysm, Abdominal / diagnostic imaging*
  • Aortography / methods
  • Asymptomatic Diseases
  • Denmark
  • Female
  • Humans
  • Image Interpretation, Computer-Assisted
  • Imaging, Three-Dimensional*
  • Male
  • Observer Variation
  • Predictive Value of Tests
  • Prognosis
  • Prospective Studies
  • Reproducibility of Results
  • Tomography, X-Ray Computed
  • Ultrasonography