Evaluation of a clinical tool to test and adjust the programmed overnight basal profiles for insulin pump therapy: a pilot study

Can J Diabetes. 2015 Oct;39(5):364-72. doi: 10.1016/j.jcjd.2015.01.005. Epub 2015 Mar 28.

Abstract

Objective: Clinical protocols for basal rate testing and adjustment are needed for effective insulin pump therapy. We evaluated the effects of a continuous glucose monitoring (CGM)-based semiautomated basal algorithm on glycemia.

Methods: We developed and piloted a basal rate analyzer that interpreted CGM data from overnight fasts and recommended dose changes for subsequent nights. Subjects uploaded data online using sensor-augmented pumps for evaluation by the analyzer after each of 5 overnight fasts conducted over 2 to 8 weeks. It was designed to be conservative and iterative, making changes that did not exceed 10% at each iteration. The standard deviation and interquartile range of CGM values from midnight to 7 am (SD12-7am and IQR12-7am) over 3 baseline and 3 postintervention nights, hypoglycemia incidence (CGM values <4.0 mmol/L), and glycated hemoglobin (A1C) were compared.

Results: Twenty subjects with mean ages of 38±13 years and A1C 7.6%±0.8% (60±8.7 mmol/mol) underwent the 5 iterations of basal assessments over 5±3 weeks. SD12-7am and IQR12-7am did not change from baseline to postintervention (1.57±0.8 to 1.63±0.8 mmol/L; p=0.35; 3.66±2.07 to 3.47±2.26 mmol/L; p=0.90). However, mean glucose values were lower between 2 to 3 am at baseline compared to postintervention; 3-night hypoglycemia incidence declined from 1.6±1.8 to 0.5±0.7 episodes (p=0.01), and A1C improved from 7.6%±0.8% to 7.4%±0.9% (60%±8.7% to 57%±9.8% mmol/mol; p=0.03).

Conclusions: The use of a basal rate analyzer was associated with reduced hypoglycemia and improved A1C. However, overnight glycemic stability was not improved. Further research into the efficacy of the CGM-based semiautomated algorithm is warranted.

Keywords: basal insulin assessment; continuous glucose monitoring; continuous subcutaneous insulin infusion; diabète de type 1; estimation de l’insuline basale; insulin pump therapy; perfusion sous-cutanée continue d’insuline; surveillance de la glycémie en continu; thérapie par pompe à insuline; type 1 diabetes.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Blood Glucose*
  • Female
  • Humans
  • Insulin Infusion Systems*
  • Male
  • Middle Aged
  • Monitoring, Physiologic / instrumentation
  • Monitoring, Physiologic / methods*
  • Pilot Projects