Clinical and radiological outcomes with the Durom™ acetabular cup for large-diameter total hip arthroplasty: 177 implants after a mean of 80 months

Orthop Traumatol Surg Res. 2015 Jun;101(4):437-41. doi: 10.1016/j.otsr.2015.02.008. Epub 2015 Apr 18.

Abstract

Background: Large-diameter metal-on-metal hip prostheses are no longer used, but their outcomes after more than 5 years are unknown. We conducted a retrospective study with a 6.8-year mean follow-up to assess clinical outcomes after Durom™ cup implantation, including the dislocation rate, comparatively to the reference metal-on-polyethylene bearing. We determined the rate of failure ascribable to Durom™ cup use. We also looked for a sharp drop in the implant survival curve during the follow-up period and for factors associated with adverse reactions to metal debris (ARMDs).

Hypothesis: We hypothesised that clinical outcomes after Durom™ cup implantation were similar to those seen with a metal-on-polyethylene bearing, except for a lower rate of dislocation.

Patients and methods: We included 177 consecutive THA procedures that were performed between 2005 and 2008 in 165 patients with a mean age of 57.6 ± 9.4 years (range, 31-76 years) and involved the implantation of a Durom™ cup, a femoral head greater than 36mm in diameter, and a PF(®) femoral stem (Zimmer, Etupes, France). The mini-posterior approach was used, with 2mm of acetabular overreaming in 82% of cases, a short femoral neck in 75% of cases, and a mean cup inclination of 34 ± 5° (range, 21-50°).

Results: Outcomes were assessed for 156 THA procedures in 146 patients after a mean follow-up of 6 years 8 months. The mean Postel-Merle d'Aubigné score improved from 9.7 ± 2.7 (range, 4-14) to 17.4 ± 1.7 (range, 15-18) and the mean Harris hip score from 45.2 ± 15.3 (range, 9-83) to 96.3 ± 7 (75-100). No episodes of dislocation were recorded. We identified 7 failures ascribable to the Durom™ cup including 6 due to ARMD and 1 to aseptic loosening. Implant survival after a mean of 80months was 95.5% (95% CI, 93.1-99.2), with no sharp drop in the survival curve.

Conclusion: The Durom™ cup eliminates the risk of hip dislocation and produces similar functional outcomes to those seen with metal-on-polyethylene bearings after a mean follow-up of 80 months. Nevertheless, given the difficulty in predicting ARMD and hypersensitivity reactions, the Durom™ cup has been discarded and patients carrying it are monitored closely.

Keywords: Durom™; Large-diameter head; Metal-on-metal; Total hip arthroplasty.

MeSH terms

  • Acetabulum / diagnostic imaging
  • Acetabulum / surgery*
  • Adult
  • Aged
  • Arthroplasty, Replacement, Hip / instrumentation*
  • Female
  • Femur / diagnostic imaging
  • Femur / surgery*
  • Follow-Up Studies
  • France
  • Hip Dislocation / diagnostic imaging
  • Hip Dislocation / surgery*
  • Hip Prosthesis / statistics & numerical data*
  • Humans
  • Male
  • Middle Aged
  • Polyethylenes*
  • Prosthesis Design
  • Radiography
  • Retrospective Studies
  • Time Factors

Substances

  • Polyethylenes