We report on a method of measuring the high-field susceptibilities of paramagnetic and diamagnetic materials using only a standard NMR system equipped with pulsed field gradients. We demonstrate the accuracy and sensitivity of the technique by measuring a series of 99.9% copper wires with diameters between 0.16 mm and 0.79 mm. We measured the volumetric susceptibility of the copper to be χ=-9.5±0.2·10(-6), which agrees with the literature value of pure copper, -9.6·10(-6). In addition to making quantitative measurements, this technique can also be used to evaluate the effectiveness of compensation schemes used to produce "zero-susceptibility" materials needed for construction of high-resolution NMR probes.
Keywords: Compensated materials; NMR; Susceptibility.
Copyright © 2015 Elsevier Inc. All rights reserved.