Aim: To evaluate the diagnostic image quality and radiation dose of low-dose 70 kV computed tomography (CT) of the paranasal sinus in comparison to 100 and 120 kV CT.
Materials and methods: CT of the paranasal sinus was performed in 127 patients divided into three groups using different tube voltages and currents (70 kV/75 mAs, ultra-low dose protocol, n = 44; 100 kV/40 mAs, standard low-dose protocol, n = 42; 120 kV/40 mAs, standard protocol, n = 41). CT dose index (CTDIvol), dose-length product (DLP), attenuation, image noise and signal-to-noise ratio (SNR) were compared between the groups using Wilcoxon-Mann-Whitney U-test. Subjective diagnostic image quality was compared by using a five-point scale (1 = non-diagnostic, 5 = excellent, read by two readers in consensus) and Cohen's weighted kappa analysis for interobserver agreement.
Results: Radiation dose was significantly lower with 70 kV acquisition than 100 and 120 kV (DLP: 31 versus 52 versus 82 mGy·cm; CTDI 2.33 versus 3.95 versus 6.31 mGy, all p < 0.05). Mean SNR (70 kV: 0.37; 100 kV: 0.21; 120 kV: 0.13; p < 0.05) and organ attenuation increased significantly with lower voltages. All examinations showed diagnostic image quality. Subjective diagnostic image quality was higher with standard protocols than the 70 kV protocol (120 kV: 5.0; 100 kV: 4.5; 70 kV: 3.5, p < 0.05) without significant differences with substantial interobserver agreement (κ > 0.59).
Conclusion: The ultra-low dose (70 kV) CT imaging of the paranasal sinus allowed for significant dose reduction by 61% and an increased attenuation of organ structures in comparison to standard acquisition while maintaining diagnostic image quality with a slight reduction in subjective image quality.
Copyright © 2015 The Royal College of Radiologists. Published by Elsevier Ltd. All rights reserved.