Importance: The Lariat device has received US Food and Drug Administration (FDA) 510(k) clearance for soft-tissue approximation and is being widely used off-label for left atrial appendage (LAA) exclusion. A comprehensive analysis of safety and effectiveness has not been reported.
Objectives: To perform a systematic review of published literature to assess safety and procedural success, defined as successful closure of the LAA during the index procedure, of the Lariat device. We performed a formal analytic review of the FDA MAUDE (Manufacturer and User Facility Device Experience) database to compile adverse event reports from real-world practice with the Lariat.
Data sources: For the systematic review, PubMed, EMBASE, CINAHL, and the Cochrane Library were searched from January 2007 through August 2014 to identify all studies reporting use of the Lariat device in 3 or more patients. The FDA MAUDE database was queried for adverse events reports related to Lariat use.
Data extractions and synthesis: Data were abstracted in duplicate by 2 physician reviewers. Events from published literature were pooled using a generic inverse variance weighting with a random effects model. Cumulative and individual adverse events were also reported using the FDA MAUDE data set.
Main outcomes and measures: Procedural adverse events and procedural success.
Results: In the systematic review, 5 reports of Lariat device use in 309 participants were identified. Specific complications weighted for inverse of variance of individual studies were urgent need for cardiac surgery (2.3%; 7 of 309 procedures) and death (0.3%; 1 of 309 procedures). Procedural success was 90.3% (279 of 309 procedures). In the FDA MAUDE database, there were 35 unique reports of adverse events with use of the Lariat device. Among these, we identified 5 adverse event reports that noted pericardial effusion and death and an additional 23 reported urgent cardiac surgery without mention of death.
Conclusions and relevance: This review of published reports and case reports identified risks of adverse events with off-label use of the Lariat device for LAA exclusion. Formal, controlled investigations into the safety and efficacy of the device for this indication are warranted.