Purpose: Dipole antennas in ultrahigh field MRI have demonstrated advantages over more conventional designs. In this study, the fractionated dipole antenna is presented: a dipole where the legs are split into segments that are interconnected by capacitors or inductors.
Methods: A parameter study has been performed on dipole antenna length using numerical simulations. A subsequent simulation study investigates the optimal intersegment capacitor/inductor value. The resulting optimal design has been constructed and compared to a previous design, the single-side adapted dipole (SSAD) by simulations and measurements. An array of eight elements has been constructed for prostate imaging on four subjects (body mass index 20-27.5) using 8 × 2 kW amplifiers.
Results: For prostate imaging at 7T, lowest peak local specific-absorption rate (SAR) levels are achieved if the antenna is 30 cm or longer. A fractionated dipole antenna design with inductors between segments has been chosen to achieve even lower SAR levels and more homogeneous receive sensitivities.
Conclusion: With the new design, good quality prostate images are acquired. SAR levels are reduced by 41% to 63% in comparison to the SSAD. Coupling levels are moderate (average nearest neighbor: -14.6 dB) for each subject and prostate B1+ levels range from 12 to 18 μT.
Keywords: Body imaging; Dipole antenna; Prostate imaging; Transceive array; Ultrahigh field.
© 2015 Wiley Periodicals, Inc.