Precision Error in Dual-Energy X-Ray Absorptiometry Body Composition Measurements in Elite Male Rugby League Players

J Clin Densitom. 2015 Oct-Dec;18(4):546-50. doi: 10.1016/j.jocd.2015.04.008. Epub 2015 Jun 10.

Abstract

Body composition analysis using dual-energy X-ray absorptiometry (DXA) is becoming increasingly popular in both clinical and sports science settings. Obesity, characterized by high fat mass (FM), is associated with larger precision errors; however, precision error for athletic groups with high levels of lean mass (LM) are unclear. Total (TB) and regional (limbs and trunk) body composition were determined from 2 consecutive total body scans (GE Lunar iDXA) with re-positioning in 45 elite male rugby league players (age: 21.8 ± 5.4 yr; body mass index: 27.8 ± 2.5 kg m(-1)). The root mean squared standard deviation (percentage co-efficient of variation) were TB bone mineral content: 24g (1.7%), TB LM: 321 g (1.6%), and TB FM: 280 g (2.3%). Regional precision values were superior for measurements of bone mineral content: 4.7-16.3 g (1.7-2.1%) and LM: 137-402 g (2.0-2.4%), than for FM: 63-299 g (3.1-4.1%). Precision error of DXA body composition measurements in elite male rugby players is higher than those reported elsewhere for normal adult populations and similar to those reported in those who are obese. It is advised that caution is applied when interpreting longitudinal DXA-derived body composition measurements in male rugby players and population-specific least significant change should be adopted.

Keywords: Bone mass; DXA; fat mass; lean mass; reproducibility.

MeSH terms

  • Absorptiometry, Photon*
  • Adolescent
  • Body Composition*
  • Bone Density
  • Football*
  • Humans
  • Male
  • Reproducibility of Results
  • Young Adult