Comparison of optimised endovaginal vs external array coil T2-weighted and diffusion-weighted imaging techniques for detecting suspected early stage (IA/IB1) uterine cervical cancer

Eur Radiol. 2016 Apr;26(4):941-50. doi: 10.1007/s00330-015-3899-5. Epub 2015 Jul 11.

Abstract

Objective: To compare sensitivity and specificity of endovaginal versus external-array coil T2-W and T2-W + DWI for detecting and staging small cervical tumours.

Methods: Optimised endovaginal and external array coil MRI at 3.0-T was done prospectively in 48 consecutive patients with stage Ia/Ib1 cervical cancer. Sensitivity/specificity for detecting tumour and parametrial extension against histopathology for a reading radiologist were determined on coronal T2-W and T2W + DW images. An independent radiologist also scored T2-W images without and with addition of DWI for the external-array and endovaginal coils on separate occasions >2 weeks apart. Cohen's kappa assessed inter- and intra-observer agreement.

Results: Median tumour volume in 19/38 cases positive on subsequent histology was 1.75 cm(3). Sensitivity, specificity, PPV, NPV were: reading radiologist 91.3 %, 89.5 %, 91.3 %, 89.5 %, respectively; independent radiologist T2-W 82.6 %, 73.7 %, 79.1 %, 77.8 % for endovaginal, 73.9 %, 89.5 %, 89.5 %, 73.9 % for external-array coil. Adding DWI improved sensitivity and specificity of endovaginal imaging (78.2 %, 89.5 %); adding DWI to external-array imaging improved specificity (94.7 %) but reduced sensitivity (66.7 %). Inter- and intra-observer agreement on T2-W + DWI was good (kappa = 0.67 and 0.62, respectively).

Conclusion: Endovaginal coil T2-W MRI is more sensitive than external-array coil for detecting tumours <2 cm(3); adding DWI improves specificity of endovaginal imaging but reduces sensitivity of external-array imaging.

Key points: • Endovaginal more accurate than external-array T2-W MRI for detecting small cervical cancers. • Addition of DWI improves sensitivity and specificity of endovaginal T2-W imaging. • Addition of DWI substantially reduces sensitivity of external-array T2-W imaging.

Keywords: Cervical cancer; Diffusion-weighted MRI; Endovaginal coil; External array coil; Fertility.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Diffusion Magnetic Resonance Imaging / methods
  • Early Detection of Cancer / methods
  • Female
  • Humans
  • Middle Aged
  • Neoplasm Staging
  • Observer Variation
  • Prospective Studies
  • Sensitivity and Specificity
  • Tumor Burden
  • Uterine Cervical Neoplasms / diagnosis*