Purpose: The purpose of this study was to evaluate the impact of offline rotational corrections and assess intrafraction motion for head and neck (H&N) cancer patients immobilized with and without a custom neck cushion.
Methods and materials: Fifty H&N cancer patients were immobilized and imaged with pretreatment and posttreatment cone beam computed tomography (CBCT) for each treatment fraction. Of these patients, 25 had a custom neck cushion added to their immobilization. Each CBCT was registered to the simulation computed tomography offline. Registrations were performed with automatching tools and a matching volume of interest that consisted of a 5-mm expansion around the mandible, occipital bone, C1/C2, and C7/T1. To determine positioning accuracy, the registration was inspected to confirm these bony anatomy structures were contained within a 3- or 5-mm expansion of the simulation position. If not, the registration was repeated with rotational corrections included and re-evaluated. For each fraction, intrafraction motion was also quantified through the difference between the pretreatment and posttreatment CBCT registration coordinates.
Results: For translational registrations, the bony anatomy in pretreatment imaging was outside the 3-mm or 5-mm expansion structure, respectively, for 49% and 15% of fractions on average for patients without a custom headrest and for 48% and 13% of fractions on average for patients with a custom headrest. The addition of rotational corrections reduced these numbers to 21% and 4% and to 28% and 6%, respectively. Intrafraction motion was significantly lower for patients immobilized with the addition of a custom neck cushion: 1.0 ± 0.5 mm compared with 1.8 ± 1.6 mm for patients with the standard headrest only (P = .02). This was reflected in posttreatment positioning accuracy, which was significantly reduced in the case of the standard headrest compared with pretreatment imaging (P values of < .001 to .048).
Conclusions: Rotational corrections significantly improved pretreatment patient positioning accuracy (P < .001). Intrafraction motion was reduced significantly through the addition of a custom neck cushion and resulted in an increase in posttreatment positioning accuracy for these patients.
Copyright © 2015 American Society for Radiation Oncology. Published by Elsevier Inc. All rights reserved.