Simulation and experimental determination of the online separation of blood components with the help of microfluidic cascading spirals

Biomicrofluidics. 2015 Jul 29;9(4):044110. doi: 10.1063/1.4927649. eCollection 2015 Jul.

Abstract

Microfluidic spirals were used to successfully separate rare solid components from unpretreated human whole blood samples. The measured separation ratio of the spirals is the factor by which the concentration of the rare component is increased due to the Dean effect present in a flow profile in a curved duct. Different rates of dilution of the blood samples with a phosphate-buffered solution were investigated. The diameters of the spherical particles to separate ranged from 2 μm to 18 μm. It was found that diluting the blood to 20% is optimal leading to a separation ratio up to 1.97. Using two spirals continuously placed in a row led to an increase in separation efficacy in samples consisting of phosphate-buffered solution only from 1.86 to 3.79. Numerical investigations were carried out to display the flow profiles of Newtonian water samples and the shear-thinning blood samples in the cross-section of the experimentally handled channels. A macroscopic difference in velocity between the two rheologically different fluids could not be found. The macroscopic Dean flow is equally present and useful to help particles migrate to certain equilibrium positions in blood as well as lower viscous Newtonian fluids. The investigations highlight the potential for using highly concentrated, very heterogeneous, and non-Newtonian fluidic systems in known microsystems for screening applications.