Objectives: Management of degenerated aortic valve bioprosthesis classically requires redo surgery, but transcatheter aortic valve-in-valve implantation is becoming a valid alternative in selected cases. In the case of a degenerated Mitroflow bioprosthesis, TAVR is associated with an additional challenge due to a specific risk of coronary occlusion. We aimed to assess the safety and feasibility of transfemoral valve-in-valve implantation of the new Edwards Sapien 3 (Edwards Lifesciences) in a degenerated Mitroflow bioprosthesis (Sorin Group, Inc).
Methods: We report here the safety and feasibility of transfemoral valve-in-valve implantation of a 23 mm Edwards Sapien 3 in a degenerated 25 mm Mitroflow valve and describe the specific assessment of the risk of coronary obstruction using a multi-imaging modality.
Results: The final result showed an absence of aortic regurgitation and a mean transvalvular gradient of 14 mm Hg. The patient had no major adverse cardiovascular events at 30-day follow-up.
Conclusion: Transcatheter valve-in-valve implantation of an Edwards Sapien 3 in a degenerated Mitroflow is feasible and safe, considering a careful assessment of the risk of coronary obstruction with Mitroflow bioprosthesis due to leaflets mounted externally to the stent.