Safety and utility of total percutaneous endovascular aortic repair with a single Perclose ProGlide closure device

J Vasc Surg. 2016 Mar;63(3):585-8. doi: 10.1016/j.jvs.2015.08.111. Epub 2015 Oct 23.

Abstract

Objective: This study evaluated the safety and efficacy of total percutaneous endovascular aortic aneurysm repair (PEVAR) with a single Perclose ProGlide device (Abbot Vascular, Santa Clara, Calif) compared with endovascular aortic repair with surgical cutdown (SEVAR).

Methods: The study included 50 abdominal aortic aneurysm patients who were treated with PEVAR with a single Perclose ProGlide device and 96 patients treated with SEVAR. Technical success was defined as successful arterial closure of the common femoral artery without the need for adjunctive surgical or endovascular procedures. The rates of complications, including bleeding requiring transfusion, infection, pseudoaneurysm, paresthesia, and lymphocele, as well as the operating room time and hospital duration were compared between the PEVAR and SEVAR groups.

Results: Technical success was obtained in all patients in the PEVAR group. One patient in the SEVAR group needed surgical repair due to access site bleeding. Complication rates were similar between the groups (4% in the PEVAR vs 8% in the SEVAR; P = .495). The PEVAR group had significantly shorter operating room times (153 ± 47 minutes vs 211 ± 88 minutes, P < .001) and hospital lengths of stay (6.7 ± 6.8 days vs 9.3 ± 4.5 days, P < .001).

Conclusions: Compared with SEVAR, PEVAR with a single ProGlide device is a safe procedure with a shorter operating room time and hospital stay, without increasing access site complications.

Publication types

  • Observational Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Aneurysm, Abdominal / diagnosis
  • Aortic Aneurysm, Abdominal / surgery*
  • Catheterization, Peripheral / adverse effects
  • Catheterization, Peripheral / methods*
  • Endovascular Procedures / adverse effects
  • Endovascular Procedures / instrumentation*
  • Equipment Design
  • Female
  • Femoral Artery*
  • Humans
  • Length of Stay
  • Male
  • Operative Time
  • Postoperative Complications / therapy
  • Punctures
  • Retrospective Studies
  • Risk Factors
  • Time Factors
  • Treatment Outcome
  • Vascular Closure Devices*