Conduction Abnormalities and Pacemaker Implantations After SAPIEN 3 Vs SAPIEN XT Prosthesis Aortic Valve Implantation

Rev Esp Cardiol (Engl Ed). 2016 Feb;69(2):141-8. doi: 10.1016/j.rec.2015.06.024. Epub 2015 Nov 2.

Abstract

Introduction and objectives: Transcatheter aortic valve implantation is increasingly used in patients with aortic stenosis. Post-procedural intraventricular conduction abnormalities and permanent pacemaker implantations remain a serious concern. Recently, the Edwards SAPIEN 3 prosthesis has replaced the SAPIEN XT. We sought to determine the incidences of new-onset intraventricular conduction abnormalities and permanent pacemaker implantations by comparing the 2 devices.

Methods: We analyzed the last consecutive 103 patients undergoing transcatheter aortic valve implantation with SAPIEN XT before SAPIEN 3 was used in the next 105 patients. To analyze permanent pacemaker implantations and new-onset intraventricular conduction abnormalities, patients with these conditions at baseline were excluded. Electrocardiograms were recorded at baseline, after the procedure, and before discharge.

Results: SAPIEN 3 was associated with higher device success (100% vs 92%; P=.005) and less paravalvular leakage (0% vs 7%; P<.001). The incidence of permanent pacemaker implantations was 12.6% (23 of 183) with no difference between the 2 groups (SAPIEN 3: 12.5% [12 of 96] vs SAPIEN XT: 12.6% [11 of 87]; P=.99). SAPIEN 3 was associated with a higher rate of new-onset intraventricular conduction abnormalities (49% vs 27%; P=.007) due to a higher rate of fascicular blocks (17% vs 5%; P=.021). There was no statistically significant difference in transient (29% [20 of 69] vs persistent 19% [12 of 64]; P=.168) left bundle branch blocks (28% [19 of 69] vs 17% [11 of 64]; P=.154) when SAPIEN 3 was compared with SAPIEN XT.

Conclusions: We found a trend toward a higher rate of new-onset intraventricular conduction abnormalities with SAPIEN 3 compared with SAPIEN XT, although this did not result in a higher permanent pacemaker implantation rate.

Keywords: Conduction abnormalities; Edwards SAPIEN 3; Edwards SAPIEN XT; Implante percutáneo de válvula aórtica; Marcapasos; Pacemaker; Transcatheter aortic valve implantation; Trastornos de la conducción.

Publication types

  • Comparative Study

MeSH terms

  • Aged, 80 and over
  • Aortic Valve / surgery
  • Aortic Valve Stenosis / diagnostic imaging
  • Aortic Valve Stenosis / surgery*
  • Arrhythmias, Cardiac / etiology*
  • Arrhythmias, Cardiac / therapy
  • Electrocardiography
  • Female
  • Heart Valve Prosthesis
  • Humans
  • Male
  • Multidetector Computed Tomography
  • Pacemaker, Artificial / statistics & numerical data*
  • Postoperative Complications / diagnostic imaging
  • Postoperative Complications / etiology
  • Prospective Studies
  • Transcatheter Aortic Valve Replacement / adverse effects*
  • Treatment Outcome