Purpose: To evaluate intraocular pressure (IOP) changes during experimental vitrectomy and the efficacy of Constellation Vision System's IOP control (IOPc) feature in reestablishing baseline pressure.
Methods: Using a pressure transducer in freshly enucleated porcine eyes, a broad range of parameters (baseline pressures, aspiration levels, and cut rates) were tested with 23- and 25-gauge probes and IOPc turned ON versus OFF.
Results: IOPc turned ON was significantly more effective than IOPc turned OFF in controlling IOP drop and stabilizing pressure during vitrectomy using a wide range of baseline pressures (20-70 mmHg). The 23-gauge system consistently presented a reduced drop from baseline compared with the 25-gauge system. The overall average drop for the 23- and 25-gauge systems was 12.79 mmHg and 21.17 mmHg, respectively. Both gauge sizes reestablished baseline pressure approximately 1.6 seconds after the initial pressure drop generated at the beginning of aspiration. A peak of IOP (overshooting) was observed when the pressure was returning to baseline using both 23- and 25-gauge systems.
Conclusion: Using IOPc feature turned ON, 23- and 25-gauge probes were effective in reestablishing and sustaining baseline infusion pressures, although 23-gauge probes showed less IOP fluctuation than did 25-gauge probes.