Accurate quantification of atherosclerotic plaque volume by 3D vascular ultrasound using the volumetric linear array method

Atherosclerosis. 2016 May:248:230-7. doi: 10.1016/j.atherosclerosis.2016.03.002. Epub 2016 Mar 15.

Abstract

Introduction: Direct quantification of atherosclerotic plaque volume by three-dimensional vascular ultrasound (3DVUS) is more reproducible than 2DUS-based three-dimensional (2D/3D) techniques that generate pseudo-3D volumes from summed 2D plaque areas; however, its accuracy has not been reported. We aimed to determine 3DVUS accuracy for plaque volume measurement with special emphasis on small plaques (a hallmark of early atherosclerosis).

Methods: The in vitro study consisted of nine phantoms of different volumes (small and medium-large) embedded at variable distances from the surface (superficial vs. >5 cm-depth) and comparison of 3DVUS data generated using a novel volumetric-linear array method with the real phantom volumes. The in vivo study was undertaken in a rabbit model of atherosclerosis in which 3DVUS and 2D/3D volume measurements were correlated against gold-standard histological measurements.

Results: In the in vitro setting, there was a strong correlation between 3DVUS measures and real phantom volume both for small (3.0-64.5 mm(3) size) and medium-large (91.1-965.5 mm(3) size) phantoms embedded superficially, with intraclass correlation coefficients (ICC) of 0.99 and 0.98, respectively; conversely, when phantoms were placed at >5 cm, the correlation was only moderate (ICC = 0.67). In the in vivo setting there was strong correlation between 3DVUS-measured plaque volumes and the histological gold-standard (ICC = 0.99 [4.02-92.5 mm(3) size]). Conversely, the correlation between 2D/3D values and the histological gold standard (sum of plaque areas) was weaker (ICC = 0.87 [49-520 mm(2) size]), with large dispersion of the differences between measurements in Bland-Altman plots (mean error, 79.2 mm(2)).

Conclusions: 3DVUS using the volumetric-linear array method accurately measures plaque volumes, including those of small plaques. Measurements are more accurate for superficial arterial territories than for deep territories.

Keywords: Accuracy; Atherosclerotic plaque; Gold-standard; Plaque volume; Three-dimensional; Vascular ultrasound.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Cohort Studies
  • Humans
  • Image Processing, Computer-Assisted / methods*
  • Imaging, Three-Dimensional
  • Linear Models
  • Phantoms, Imaging
  • Plaque, Atherosclerotic / diagnostic imaging*
  • Rabbits
  • Reproducibility of Results
  • Software
  • Ultrasonography*