Microbore liquid chromatography and refractive index gradient detection of low-nanogram and low-ppm quantities of carbohydrates

J Chromatogr. 1989 Feb 17;464(1):83-91. doi: 10.1016/s0021-9673(00)94225-0.

Abstract

A refractive index gradient detector is presented as a universal detector in the microbore high-performance liquid chromatography analysis of carbohydrates. Simultaneously, low-ng and low-ppm injected quantities of carbohydrates were detected at the 3 x root-mean-square baseline noise level. A typical microbore high-performance liquid chromatography chromatogram separating fructose from sucrose followed by refractive index gradient (RIG) detection is reported. Use of a position sensitive detector (PSD) in the RIG detector design is reported and experimental considerations discussed. Optimization of the PSD-based RIG detector is addressed. Potential for the device in industrial and clinical applications is considered.

Publication types

  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Carbohydrates / analysis*
  • Chromatography, High Pressure Liquid
  • Refractometry