Local lymph node assay: 5-bromo-2-deoxyuridine-flow cytometry method (LLNA: BrdU-FCM) is a modified non-radioisotopic technique with the additional advantages of accommodating multiple endpoints with the introduction of FCM, and refinement and reduction of animal use by using a sophisticated prescreening scheme. Reliability and accuracy of the LLNA: BrdU-FCM was determined according to OECD Test Guideline (TG) No. 429 (Skin Sensitization: Local Lymph Node Assay) performance standards (PS), with the participation of four laboratories. Transferability was demonstrated through successfully producing stimulation index (SI) values for 25% hexyl cinnamic aldehyde (HCA) consistently greater than 3, a predetermined threshold, by all participating laboratories. Within- and between-laboratory reproducibility was shown using HCA and 2,4-dinitrochlorobenzene, in which EC2.7 values (the estimated concentrations eliciting an SI of 2.7, the threshold for LLNA: BrdU-FCM) fell consistently within the acceptance ranges, 0.025-0.1% and 5-20%, respectively. Predictive capacity was tested using the final protocol version 1.3 for the 18 reference chemicals listed in OECD TG 429, of which results showed 84.6% sensitivity, 100% specificity, and 88.9% accuracy compared with the original LLNA. The data presented are considered to meet the performance criteria for the PS, and its predictive capacity was also sufficiently validated.
Keywords: Flow cytometry; LLNA: BrdU-FCM; Local lymph node assay; Performance standards; Skin sensitization; Validation study.