Aims: High-energy resolution and sensitivity of novel cadmium-zinc-telluride (CZT) detector equipped SPECT systems facilitate simultaneous imaging of multiple isotopes and may enhance the detection of molecular/cellular signals. This may refine the detection of endocarditis. This study was designed to determine the feasibility and diagnostic accuracy of simultaneous imaging of inflammation with 111In-labeled white blood cells (WBCs) and myocardial perfusion with 99mTc-sestamibi, for localization of WBCs relative to the valve plane in suspected endocarditis.
Methods and results: A dedicated cardiac CZT camera (Discovery 530c, GE Healthcare) was employed. Anthropomorphic thorax phantom studies were followed by clinical studies in 34 patients with suspected infection of native valves (n = 12) or implants (n = 22). Simultaneous 111In-WBC/99mTc perfusion imaging was performed, and compared with standard 111In-WBC planar scintigraphy and SPECT-CT. Phantom studies ruled out significant radioisotope crosstalk. Downscatter on 99mTc images was not observed for 111In activity as high as 2.5*99mTc activity. In patients, image quality was superior for CZT imaging vs. conventional SPECT-CT and planar scintigraphy (P < 0.01). Cadmium-zinc-telluride dual isotope imaging improved reader confidence for detection of inflammatory foci. Diagnostic accuracy based on surgery or Duke Criteria during follow-up was highest for CZT imaging (P < 0.001).
Conclusion: Novel CZT SPECT technology improves the accuracy of molecular/cellular cardiac imaging. Simultaneous multi-isotope imaging with 111In and 99mTc is feasible and aids in the workup of suspected endocarditis.
Keywords: Endocarditis; Single-photon emission computed tomography; Solid-state detector imaging; White blood cell tracking.
Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2016. For permissions please email: [email protected].